9. Fieldbus interface | 9.2 - Scale protocol | 118
Field
Function
ADUError
AD conversion error (OR function of bits E1, E3, E7).
AboveMax
The weight value has exceeded the Max (FSD), but is still
within Max + permissible overload (gross ≤Max + over
load).
Overload
The weight value has exceeded the measuring range. No
valid weight data is specified (gross >Max + overload);
ERR 2
.
BelowZero
The weight value is negative (gross < 0d).
CenterZero
The weight value is within center zero (0 ± 0.25 d)
InsideZSR
The gross weight value is within the zero setting range.
Standstill
The scale is stable.
OutOfRange
Below zero or above Max (FSD).
E9
The measuring signal is higher than the permissible range
of 36 mV. Cannot read weight values from ADC (ana
log-digital converter) (
ERR 9
).
E7
The measuring signal is negative (inverse conversion) (
ERR
7
)
E6
Sense voltage not present or too low (
ERR 6
)
E3
The measuring signal is >36 mV (no end of conversion)
(
ERR 3
)
CmdError
Error during execution (CmdError); e.g., the "taring" oper
ation is not processed, because the scale is not at a stand
still. The error is stored in
LastError
(function number 4).
The bit is reset with the
ResetError
bit (function num
ber 121, see Chapter
Function number 112–121: transi
tion-controlled action bits (write)
).
ActionActive
The device is busy executing a function (e.g., waiting for
downtime for taring).
PowerFail
Power failure; is always set after power on. The
PowerFail
bit is reset with the
ResetPWF
bit (function number 85, see