ST2826 Operation manual
Ver1.0
◇
99
and limit comparison.
Command hierarchical chart:
LIST :FREQuency <sweep point>[,<sweep point> *]
:VOLTage <sweep point>[,<sweep point> *]
:CURRent <sweep point>[,<sweep point> *]
:BIAS :VOLTage <sweep point>[,<sweep point> *]
:CURRent <sweep point>[,<sweep point> *]
:MODE SEQuence
STEPped
:BAND<n> A[,<low limit n>,<high limit n>]
B
OFF
:FREQuency
is used to clear the previous sweep point and set the frequency of the sweep
point.
command syntax
:
LIST:FREQuency <value> [,<value> *]
note
:
‖ *‖ shows that the sweep points are 10 points at most.
<value> : NR1, NR2 or NR3 format
<value>: must be in the range of 20HZ to 300KHZ
(
20HZ-200KHZ for ST2826
)
, or
return ―error‖.
e.g.
:
WrtCmd(―LIST:FREQ 1E3,2E3,3E3,4E3‖);set the frequency of sweep point to 1KHZ
;
;set the frequency of sweep point to 2KHZ
;
;set the frequency of sweep point to 3KHZ
;
;set the frequency of sweep point to 4KHZ
note
:
HZ
(
hertz
)
is postfix unit, MAHZ and MHZ are MHz(1E6 Hz)
。
query syntax
:
LIST:FREQuency?
return
:
<NR3>[,<NR3> *]<NL^END>
:
VOLTage
is used to clear the previous list sweep level and set the sweep leve.
command syntax
:
LIST:VOLTage <value>[,<value> *]
note
:
‖ *‖ shows that the sweep points are 10 points at most.