ST2826 Operation manual
Ver1.0
◇
110
0
char ‗1‘(49):enable the short correction, which is equivalent to ON.
char ‗0‘(48):forbid the short correction, which is equivalent to OFF.
e.g.
:
WrtCmd(―CORR:SHOR:STAT ON‖)
query syntax
:
CORRection:SHORt:STATe?
return
:
<NR1><NL^END>
:LOAD:STATe
is used to set the load correction function. Char ‗?‘ is used to query the state of
the load correction.
command syntax
:
ON
CORRection:LOAD:STATe OFF
1
0
char ‗1‘(49):enable the load correction, which is equivalent to ON.
char ‗0‘(48):forbid the load correction, which is equivalent to OFF.
e.g.
:
WrtCmd(―CORR
:
LOAD
:
STAT ON‖)
query syntax
:
CORRection:LOAD:STATe?
return
:
<NR1><NL^END>
:LOAD:TYPE
is used to set the combination parameter function for the load correction. Char
‗?‘is used to query the type of the combination parameter.
Fuction
:
CPD: set the function to Cp-D
LPRP: set the function to Lp-Rp
CPQ: set the function to Cp-Q
LSD : set the function to Ls-D
CPG: set the function to Cp-G
LSQ: set the function to Ls-Q
CPRP: set the function to Cp-Rp
LSRS: set the function to Ls-Rs
CSD: set the function to Cs-D
RX: set the function to R-X
CSQ: set the function to Cs-Q
ZTD : set the function to Z-θ°
CSRS: set the function to Cs-Rs
ZTR : set the function to Z-θr
LPQ: set the function to Lp-Q
GB: set the function to G-B
LPD : set the function to Lp-D
YTD: set the function to Y-θ°
LPG: set the function to Lp-G
YTR: set the function to Y-θr
e.g.
:
WrtCmd(―CORR:LOAD:TYPE CPD‖)
Summary of Contents for ST2826
Page 2: ......
Page 10: ......
Page 51: ...ST2826 Operation manual Ver1 0 41 Figure 3 15 ...
Page 61: ...ST2826 Operation manual Ver1 0 51 ...
Page 144: ...ST2826 Operation manual Ver1 0 134 Figure 9 8 the input circuit for control signal ...