Commands listed in alphabetical order
Returns
The query response is returned in high precision NR3 format (up to 12 digits with
more than 4 trailing 0 digits after the decimal point is omitted).
Examples
AFG:PERIOD 1
sets the AFG period value to 1 second.
AFG:PERIOD
? might return :AFG:PERIOD 3.2000E-6 indicating the AFG period
is set to 3.2
μ
s.
AFG:PULse:WIDth
Sets (or queries) the AFG pulse width, in seconds.
Conditions
Requires option AFG.
Group
AFG
Syntax
AFG:PULse:WIDth <NR3>
AFG:PULse:WIDth?
Arguments
NR3
is the
fl
oating point number that represents the pulse width, in seconds.
Examples
AFG:PULSE:WIDTH 100.0E-6
sets the AFG pulse width to 100 microseconds.
AFG:PULSE:WIDTH
? might return
:AFG:PULSE:WIDTH 1.0000E-6
indicating
the pulse width is set to 1
μ
s.
AFG:RAMP:SYMmetry
Sets (or queries) the AFG ramp symmetry in percent. Minimum is 0.0%,
maximum is 100.0% and increment is 0.10%.
Conditions
Requires option AFG.
Group
AFG
Syntax
AFG:RAMP:SYMmetry <NR3>
AFG:RAMP:SYMmetry?
Arguments
Floating point number that represents the AFG ramp symmetry, as a percentage.
2-106
MSO54, MSO56, MSO58, MSO58LP Programmer
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...