Commands listed in alphabetical order
Examples
TRIGGER:A:UPPERTHRESHOLD:CH1 1.3
This command sets the A trigger level
for Channel 1 to 1.3 V level.
TRIGGER:A:UPPERTHRESHOLD:CH2?
might return
:TRIGGER:A:UPPERTHRESHOLD:CH2 1.3000E+00
, indicating that the A
trigger level for Channel 2 is set to 1.3 V.
TRIGger:{A|B}:WINdow:CROSSIng
This command sets or queries the window trigger threshold crossing of the
selected trigger source. The threshold crossing selection is only effective when
:TRIGger:{A|B}:WINdow:WHEn is INSIDEGreater or OUTSIDEGreater.
Group
Trigger
Syntax
TRIGger:{A|B}:WINdow:CROSSIng {UPPer|LOWer|EITher|NONe}
Arguments
UPPer
if
:TRIGger:{A|B}:WINdow:WHEn
is
INSIDEGreater
, the instrument
triggers when the signal remains between the upper and lower thresholds for longer
than the time limit (
:TRIGger:{A|B}:WINdow:WIDTH
) and then exits through
the upper threshold. If
:TRIGger:{A|B}:WINdow:WHEn
is
OUTSIDEGreater
,
the instrument triggers when the signal remains above the upper threshold for
longer than the time limit (
:TRIGger:{A|B}:WINdow:WIDTH
) and then crosses
downward through the upper threshold.
LOWer
if
:TRIGger:{A|B}:WINdow:WHEn
is
INSIDEGreater
, the instrument
triggers when the signal remains between the upper and lower thresholds for longer
than the time limit (
:TRIGger:{A|B}:WINdow:WIDTH
) and then exits through
the lower threshold. If
:TRIGger:{A|B}:WINdow:WHEn
is
OUTSIDEGreater
,
the instrument triggers when the signal remains below the lower threshold for
longer than the time limit (
:TRIGger:{A|B}:WINdow:WIDTH
) and then crosses
upwards through the lower threshold.
EITher
if
:TRIGger:{A|B}:WINdow:WHEn
is
INSIDEGreater
, the instrument
triggers when the signal remains between the upper and lower thresholds for
longer than the time limit (
:TRIGger:{A|B}:WINdow:WIDTH
) and then exits
through either the upper or lower threshold. If
:TRIGger:{A|B}:WINdow:WHEn
is
OUTSIDEGreater
, the instrument triggers when the signal remains either
above the upper threshold or below the lower threshold for longer than the time
limit (
:TRIGger:{A|B}:WINdow:WIDTH
) and then crosses a threshold.
NONe
if
:TRIGger:{A|B}:WINdow:WHEn
is
INSIDEGreater
, the instrument
triggers when the signal remains between the upper and lower thresholds for
longer than the time limit (
:TRIGger:{A|B}:WINdow:WIDTH
) without crossing
through either the upper or lower threshold. If
:TRIGger:{A|B}:WINdow:WHEn
is
OUTSIDEGreater
, the instrument triggers when the signal remains
2-872
MSO54, MSO56, MSO58, MSO58LP Programmer
Summary of Contents for MSO54
Page 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Page 2: ......
Page 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Page 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Page 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Page 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...