System 8700 User Manual
Rev 0.95
Page 19 of 32
4.3.1.1.
Status reporting
The status reporting is done via the OPERation status register with the command
STATus:OPERation. The different status, warnings and errors are split up in separate arrays,
see below. The array is summed up on the level above with an “or” so that if one or
more bits
are set in an array, the bit on the level above will also be set. New sub commands have been
constructed to get the individual sub arrays e.g. to read the status bits in register 1, the
command STATus:QUEStionable:REG1 is used.
In the tables below the Bit indicates what bit the signal corresponds to after the decimal
number returned by the command has been converted to a binary digit, see also the example
after Table 6 for further explanation. The Input column indicates what digital input the signal is
connected to. If no input is indicated the signal is generated internally.
4.3.1.2.
Status, bit 9
Command: STATus
Bit
Signal
Interpretation
0
Output On
1: Output is on
1
Ready
1: regulation reports output within defined limits
2
-
Not used
3
-
Not used
4
Sum interlock
1: one or more interlocks present
5
-
Not used
6
-
Not used
7
Operation Status
1: Further info in Operation Status registers available
Table 2 The status register