262
The scale does not report an error, the measurement is stable
in the second measuring range.
•
LO
– the value of
LO
threshold in calibration/adjustment unit is returned.
•
Process status
– determines the status of dosing process
byte no.
Decimal value of
a variable (Dec)
process status
B1 B0
0
process disabled (inactive)
0
0
1
dosing start
0
1
2
dosing stop
1
0
3
end of dosing process
1
1
•
Inputs status
– determines the status of inputs in a terminal PUE7:
Input
no. 12 11 10 9 8 7 6 5 4 3 2 1
Decimal value of
a variable (Dec)
OFF
0 0 0 0 0 0 0 0 0 0 0 0
0
ON
1 1 1 1 1 1 1 1 1 1 1 1
4094
•
MIN
– set value of
MIN Threshold
(in calibration/adjustment unit)
is returned.
•
MAX
– set value of
MAX Threshold
(in calibration/adjustment unit)
is returned.
•
Lot number
– response is the value of lot number (numeric range
0 - 2147483647).
•
Operator
– response is the value of a logged operator (numeric range
0 - 32767).
•
Product
– response is the value of a selected product (numeric range
0 - 32767).
•
Client
– response is the value of a code of a selected client
(numeric range 0 - 32767).
•
Packaging
– response is the value of a code of a selected packaging
(numeric range 0 - 32767).