264
•
command
– basic command. Setting of an appropriate command’s
byte is directly responsible for carrying out a task, in accordance with
below table:
command
bytes
task Dec
0 scale
zeroing
1
1 scale
tarring
2
5
dosing process start
32
6
dosing process stop (emergency stop)
64
For instance:
0000 0000 0010 0000
– command carried out, dosing process start
•
complex command
- Setting of an appropriate command’s byte
is directly responsible for carrying out a task, in accordance with
below table:
command
bytes
task Dec
0
setting tare value
1
1
setting
LO
threshold value
2
2
setting outputs status
4
3
setting
MIN threshold
value
8
4
setting
MAX threshold
value
16
For instance:
0000 0000 0000 0010
– command carried out, LO threshold set to value
given in LO parameter (address 10); See Table “Input variables”.
Notice:
A complex command requires setting an appropriate address parameter
from 6 to 28. See table “Input variables”.
•
Platform
– complex command parameter: number of a weighing
platform.
•
tare
– complex command parameter – tare value (in calibration /
adjustment unit).