Remote Commands Reference
R&S
®
RTC1000
248
User Manual 1335.7352.02 ─ 02
MASK:ACTion:WFMSave:DESTination
<File>
Defines the path and filename for a waveform data that will be saved on mask viola-
tion. The file format is CSV, the filename is incremented automatically
You can change the storage location, file name and/or file format manually in the FILE
> "Waveforms" menu. Remote control uses the recent settings.
Parameters:
<File>
String parameter
Example:
MASK:ACT:WFMS:DEST "/USB_FRONT/MASKS/VIOL"
On first violation, the waveform data is saved to
VIOL.CSV
, on
second violation to
VIOL01.CSV
, the third to
VIOL02.CSV
...
MASK:RESet:COUNter
Sets the counters of passed and failed acquisitions to zero.
Usage:
Event
13.10 Function Generator
..........................................................................................249
GENerator:FUNCtion:RAMP:POLarity
..............................................................................249
GENerator:FUNCtion:PULSe:DCYCle
..............................................................................249
.......................................................................................... 249
GENerator:FUNCtion
<Function>
Selects the generator function.
Parameters:
<Function>
DC | SINusoid | SQUare | PULSe | TRIangle | RAMP
*RST:
SIN
GENerator:VOLTage
<Amplitude>
Defines the amplitude value (peak-peak value) of the selected generator function.
Parameters:
<Amplitude>
Range:
6.0000E-02 to 6.00
*RST:
5.0000E-01
Default unit: Vpp
Function Generator