4-65
Programming Commands
CALCulate3 Subsystem
FPERot:PEAK?
Queries the peak data of the selected modes.
Syntax
:CALCulate3:FPERot:PEAK{[:WAVelength] | :FREQuency | :WNUMber | :POWer{[:DBM]|:WATTs}}?
Argument
Description
WAVelength
Returns the peak wavelength of the selected modes.
FREQuency
Returns the peak frequency of the selected modes.
WNUMber
Returns the peak wavenumber of the selected modes.
POWer
Returns the peak amplitude of the selected modes in dBm or
watts.
Example Query
Response
WAVelength
+1.54073400E–006
FREQuency
+1.94014
WNUMber
+6.49003
POWer
–9.09000
(dBm)
+1.23183800E–004
(watts)
Attribute
Summary
Preset
State: not affected
*RST State: not affected
SCPI Comp liance: instrument sp ecific
Query only