Digital Input Commands
These commands read individual bits or complete ports. Some digital I/O
configuration commands may affect how these bits are read. The port
parameter is in the form snn, where s is the slot number and nn is the
first channel on the digital port. The port parameter is plug-in module
dependent. Valid port numbers for each plug-in module are shown
beginning on page 70.
SENSe:DIGital:DATA
:BIT? <bit_port>
This query reads a bit from the specified bit_port and returns either a “0”
or a “1”. bit_port is in the form of snn, where s is the slot number and nn
is the bit number. The built-in 4-bit digital I/O port has bit numbers 091
to 094. The plug-in module bit numbers are module dependent (see page
70).
SENSe:DIGital:DATA
[:<BYTE|WORD|LWORD>][:VALue]? <port>
This query reads data from the specified port and returns an integer
representing the binary-weighted value of the bits. The value returned
depends upon the first parameter as follows:
•
BYTE -
reads 8 bits and returns an integer between 0 and 255 (00
h
and FF
h
).
•
WORD -
reads 16 bits and returns an integer between -32768 and
+32767 (-8000
h
and +7FFF
h
). Negative numbers returned are 2’s
complement numbers.
•
LWORD
- reads 32 bits and returns an integer between -2
31
and
+(2
31
-1) (-80000000
h
and 7FFFFFFF
h
). Negative numbers returned
are 2’s complement numbers.
Numbers greater than +32767 (
WORD
) and +(2
31
-1) (
LWORD
) will be
returned as negative numbers. When the returned data is negative, you
must calculate the actual bit state from the 2’s complement form.
Summary of Contents for Agilent 3499A
Page 13: ...1 1 Quick Start ...
Page 27: ...2 2 Front Panel Operation ...
Page 55: ...3 3 System Overview ...
Page 77: ...4 4 Features and Functions ...
Page 113: ...5 5 Remote Interface Reference ...
Page 164: ...164 5 ...
Page 165: ...6 6 Error Messages ...
Page 173: ...7 7 Plug in Modules ...
Page 256: ...256 Chapter 7 Plug in Modules 44475A Breadboard Module 7 ...
Page 261: ...261 Chapter 7 Plug in Modules 44476B Microwave Switch Module 4 7 The 44476B is shown below ...
Page 286: ...286 Chapter 7 Plug in Modules Terminals and Connections Information 7 ...
Page 288: ...288 7 ...
Page 289: ...8 8 Application Programs ...
Page 299: ...9 9 Specifications ...
Page 343: ...343 Chapter 9 Specifications 44475A Breadboard Module 4 9 ...