Polycold Cryochiller
Host Commands
Installation and Operation Manual
System Configuration Commands
Brooks Automation
214072 Revision B
7-13
LEGACY_REMOTE,COMP_SIGNAL_LOGIC Compressor Input Signal Logic
Purpose:
Get/Set the Legacy Remote compressor input signal logic. Default setting is
equivalent to jumper “J11” being placed on PFC system control board. The
compressor will not be controlled by Legacy Remote compressor on/off input
signal.
Monitor Data:
N/A
Parameter(s):
? or [n]
[n]
0 = Compressor controlled by Legacy Remote compressor input signal
Equivalent to “J11” jumper being removed from PFC system control board
1 = Compressor not controlled by Legacy Remote compressor input signal.
Equivalent to “J11” jumper being placed on PFC system control board
Return Value:
[A] or [A][n]
Data Type:
[n]: Integer
Default Value:
[n] = 1
Min: [n] = 0
Max: [n] = 1
Units:
N/A
Security Level
Get: 1
Set: 3
Example:
“LEGACY_REMOTE,COMP_SIGNAL_LOGIC,0” Set the Legacy Remote
interface to use the state of the compressor on/off input signal to control the
compressor.
Interface
RS-232, Ethernet