Remote Control Commands
R&S
®
SGT100A
341
User Manual 1176.8674.02 ─ 07
Example:
See
Manual operation:
See
[:SOURce]:FREQuency:OFFSet
<Offset>
Sets a frequency offset, for example include the frequency shift of downstream instru-
ment.
Note:
Enabled frequency offset affects the result of the query
The query returns the frequency, including frequency offset.
Parameters:
<Offset>
float
Range:
-3e9 to 3e9
Increment: 1E-3
*RST:
0
Example:
SOURce:FREQuency:OFFSet 0
SOURce:FREQuency:CW 2143000000
SOURce:FREQuency:OFFSet 20000000
SOURce:FREQuency:CW?
// 2163000000
Manual operation:
See
[:SOURce]:OPMode
<OpMode>
Sets the operation mode.
Parameters:
<OpMode>
NORMal | BBBYpass
NORMal
normal operation
BBBYpass
Baseband bypass mode
*RST:
NORMal
Manual operation:
See
[:SOURce]:LOSCillator:SOURce
<Source>
Selects the source of the local oscillator signal.
Parameters:
<Source>
INTernal | EXTernal
INT
: use built in oscillator;
EXT
: use signal at LO/ REF IN con-
nector
Manual operation:
See
SOURce Subsystem