TDI-Dynaload Division
WCL488 Series Water Cooled Electronic Loads
Operation & Programming Manual
402828, Rev. B1
Page 73
R1 Constant Resistance Pulsing Base Level
Description
This command is used to set the base level when pulse loading in the
Constant Resistance mode.
NOTE:
CRH or CRL must have been previously set.
Command Syntax
R1 <NR2>
Example
R1 10.0
This will set the base resistance to 10 Ohms.
Query Syntax
R1?
Returned Parameters TEXT ON: <NR2> ohms
TEXT OFF: <NR2>
R2 Constant Resistance Pulsing Peak Level
Description
This command is used to set the peak level relative to the base level
when pulse loading in the Constant Resistance mode. This resistance
will be applied in parallel to the base resistance.
NOTE:
CRH or CRL must have been previously set.
Command Syntax
R2 <NR2>
Example
R2 1.1
This will set the peak level to 1.1 Ohms that will be applied in
parallel with the base level. Therefore if the base level setting is 10
Ohms, the peaks will be 0.99 Ohm.
Query Syntax
R2?
Returned Parameters TEXT ON: <NR2> ohms
TEXT OFF: <NR2>