WCL488 Series Water Cooled Electronic Loads
Operation & Programming Manual
TDI-Dynaload Division
Page 72
402828, Rev. B1
P1 Constant Power Pulsing Base Level
Description
This command is used to set the base level when pulse loading in the
Constant Power mode.
Command Syntax
P1 <NR2>
Example
P1 25.3
This will set the base Constant Power level to 25.3 Watts.
Query Syntax
P1?
Returned Parameters TEXT ON: <NR2> watts
TEXT OFF: <NR2>
P2 Constant Power Pulsing Peak Level
Description
This command is used to set the peak level relative to the base level
when pulse loading in the Constant Power mode.
Command Syntax
P2 <NR2>
Example
P2 20.0
This will set the peak level to 20 Watts above the base level.
Therefore if the base level setting is 25.3, the peaks will be 45.3
Watts.
Query Syntax
P2?
Returned Parameters TEXT ON: <NR2> watts
TEXT OFF: <NR2>
PL Power Limit
Description
This command is used to set the Power Limit set point. The Power
Limit circuit will not allow the load power to exceed this level.
Command Syntax
PL <NR2>
Example
PL 250.0
This will set the Power Limit to 250 Watts.
Query Syntax
PL?
Returned Parameters TEXT ON: <NR2> watts
TEXT OFF: <NR2>