REMOTE CONTROL
173
CONFigure:FRESistance
Sets measurement to 4W Resistance on the first display and specifies the
range/resolution.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf> |
MIN | MAX | DEF)]
Example: CONF:FRES 10e3,MAX
Sets the measurement mode to 4W with a range of 10k
Ω
at the maximum
resolution.
CONFigure:FREQuency
Sets measurement to Frequency on the first display and specifies
range/resolution.
Parameter:
[None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf> |
MIN | MAX | DEF)]
Example: CONF:FREQ MAX,MAX
Sets the frequency measurement range to max and the resolution to max.
CONFigure:PERiod
Sets measurement to Period on the first display and specifies the
range/resolution.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf> |
MIN | MAX | DEF)]
Example: CONF:PER
Sets the DMM to period measurement using the previous
range/resolution.
CONFigure:CONTinuity
Sets measurement to Continuity on the first display.
Parameter: None
CONFigure:DIODe
Sets measurement to Diode on the first display.
Parameter: None
CONFigure:TEMPerature:TCOuple
Sets measurement to Temperature thermocouple (T-CUP) on the first
display.
Parameter: [None] | [Type(B | E | J | K | N | R | S | T)]
Example: CONF:TEMP:TCO J
Sets the measurement mode to TCO with a type J sensor.
CONFigure:TEMPerature:FRTD
Sets the measurement mode to 4W RTD measurement mode on the first
display. Sets the sensor type.
Parameter: [None] | [Type(PT100 | D100 | F100 | PT385 | PT3916 | USER)]
Example: CONF:TEMP:FRTD PT100
Sets the sensor type to PT100 and sets the measurement mode to 4W RTD