REMOTE CONTROL
185
[SENSe:]PERiod:INPutjack?
Returns the assigned input port used for the period function.
Return Parameter: VOLT, 1A, 10A
[SENSe:]DETector:BANDwidth
Sets the AC bandwidth (AC filter).
Parameter: (3 | 20 | 200)
Example: SENS:DET:BAND 20
Sets the AC bandwidth to 20Hz.
[SENSe:]DETector:BANDwidth?
Returns the AC bandwidth.
[SENSe:]ZERO:AUTO
Sets the Auto zeroing mode to on, off or once only.
Parameter: ON | OFF | ONCE
Example: SENS:ZERO:AUTO ONCE
Sets the auto zeroing to once only.
[SENSe:]ZERO:AUTO?
Returns the Auto zero mode.
Return Parameter: 0|1, 1=ON, 0=OFF
[SENSe:]GAIN:AUTO
Sets the Auto gain mode to on, off or once only.
Parameter: ON | OFF | ONCE
Example: SENS:GAIN:AUTO OFF
Turns the Auto gain mode off.
[SENSe:]GAIN:AUTO?
Returns the Auto gain mode.
Return parameter: 0|1, 1=ON, 0=OFF
[SENSe:]CONTinuity:THReshold
Sets the continuity threshold in ohms.
Parameter: <NRf> (0 ~ 1000)
Example: SENS:CONT:THR 500
Sets the continuity threshold to 500
[SENSe:]CONTinuity:THReshold?
Returns the continuity threshold.
[SENSe:]CURRent:DETect
Sets the current auto-detect mode on or off for the current functions.
Parameter: ON | OFF
Example: SENS:CURR:DET ON
Turns the current auto-detect on for the current function.