REMOTE CONTROL
179
MEASure:TEMPerature:FRTD?
Returns the 4W RTD temperature for the selected sensor type on the first
display.
Parameter:[NONE] | PT100 | D100 | F100 | PT385 | PT3916 | USER
Example: MEAS:TEMP:FRTD? PT100
>
+0.5
Returns the temperature.
MEASure:TEMPerature:RTD?
Returns the 2W RTD temperature for the selected sensor type on the first
display.
Parameter:[NONE] | PT100 | D100 | F100 | PT385 | PT3916 | USER
Example: MEAS:TEMP:RTD? PT100
>
+0.5
Returns the temperature.
MEASure2:VOLTage:DC?
Returns the DC voltage measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: MEAS2:VOLT:DC ?
>+0.488E-4
Returns the DC voltage measurement as 0.0488 mV.
MEASure2:VOLTage:AC?
Returns the AC voltage measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: MEAS2:VOLT:AC ?
>+0.511E-3
Returns the AC voltage measurement as 0.511 mV.
MEASure2:CURRent:DC?
Returns the DC current measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: MEAS2:CURR:DC ?
>+0.234E-4
Returns the DC current measurement as 0.0234 mA.
MEASure2:CURRent:AC?
Returns the AC current measurement on the second display.
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF),Resolution(<NRf>|
MIN | MAX | DEF)]
Example: MEAS2:CURR:AC ?
>
+0.387E-2
Returns the AC current measurement.