Series 3700 System Switch/Multimeter Reference Manual
Section 9
:
Instrument Control Library (ICL)
Document Number: 3700S-901-01 Rev. A / August 2007
9-149
localnode.define.*
.MAX_TIMERS
.MAX_DIO_LINES
.MAX_TSPLINK_TRIGS
.MAX_BLENDERS
.MAX_BLENDER_INPUTS
.MAX_LAN_TRIGS
Example
To read features maximum number of timers:
operreg = localnode.define.MAX_TIMERS
localnode.linefreq
Attribute
Use to read power line frequency detected at power up.
Usage
To read line frequency:
frequency = localnode.linefreq
Remarks
To achieve optimum noise rejection when performing measurements at integer
NPLC apertures, the line frequency setting must match the frequency (50Hz or
60Hz) of the AC power line.
When used in an expanded system (TSP-Link), localnode.linefreq is sent to the
Remote Master node only. Use node[N].linefreq (where N is the node number) to
send the command to any node in the system. See for details on TSP-Link.
Example
To get the line frequency:
print(localnode.linefreq)
localnode.model
Attribute
Use to query model number.
Usage
value = localnode.model
value: Represents model number.
Example
To print model number:
print(localnode.model)
Output:
3700S