7352 Series Digital Multimeter Operation Manual
6.6.4 SCPI Command Reference
6-42
System
Calibration
:CALibration:SECure:STATe
{ON,1|OFF,0}
ON, 1: Sets the calibration mode to ON.
OFF, 0: Sets the calibration mode to OFF.
(Writes the calibration factor when leaving the cali-
bration mode.)
:CALibration:SECure:STATe?
Reply: 1 or 0
:CALibration:ABORt
Cancels (sets to OFF) the calibration mode.
(Does not write the calibration factor.)
:CALibration <int>
Enters the STD value. (Displayed count value)
<int>: 0 to
999999
:CALibration:VALue <real>
Enters the STD value. (Displayed value)
<real>: Displayed value on STD
:CALibration:STRing <"str">
Sets the calibration information.
<“str”>:Enclose the character strings by using (").
Up to 50 characters (one byte alphanumeric
characters) can be stored.
:CALibration:STRing?
Reply: "xxxxx
xxx"
Setting
parameter
*SAV <int>
<int>: 0 to 3
Saves the setting parameters in area [int] of the
non-volatile memory.
*RCL <int>
<int>: 0 to 3
Loads the setting parameters from area [int] of the
non-volatile memory.
Error log
reading
:SYSTem:ERRor?
Reads the error details.
Reply:
ddd,"xxxxxxxxxx"
• Up to 20 error queues can be stored. The error is
output according to the FIFO system.
• If more than 20 errors occur, the last error is
replaced with -350, “Queue overflow.” Any
errors after this are not saved.
• If no error occurs, +000, “No error” is returned.
SCPI version
:SYSTem:VERSion?
Reads SCPI version.
Reply: 1991.0
Item
SCPI command
Description
Initial
value
Up to 50 characters
Error character strings
(Up to 80 characters)
Error code