7B
REV. C
B-10
GPIB PROGRAMMING EXAMPLE FOR THREE PHASE UNIT
1. EXAMPLE:
Setup single range BL to output 12/0 V @ 6/0 Hz and read back actual voltage and
current from phase 2 and the average voltage. Assume for this example a load is connected
drawing 1.5 amps RMS from phases 1 and 2, /0 amps from phase 3.
Transmit setup command:
FNC ACS: CH/0 SET VOLT 12/0 SET FREQ 6/0 <cr><lf>
Transmit status command to request status response:
STA <cr><lf>
Talk address the BL source to enable response:
<space><cr><lf>:
Positive response indicating status is OK.
Transmit output relay close command:
CLS :CH/0 <cr><lf>
Transmit status command to request status response:
STA <cr><lf>
Talk address the BL source to enable response:
<space><cr><lf>
Positive response indicating status is OK.
Transmit fetch volt command for phase 2:
FTH VOLT2 <cr><lf>
Talk address the bl source to enable response:
12/0./0 <cr><lf>
Transmit fetch volt command for average of 3 phases:
FTH VOLT <cr><lf>
Talk address the BL source to enable response:
12/0./0 <cr><lf>
Transmit fetch current command for phase 2:
FTH CURR2 <cr><lf>
Talk address the BL source to enable response:
1.5 <cr><lf>
Transmit FTH current command for average of 3 phases:
FTH CURR <cr><lf>
Talk address the BL source to enable response:
1./0 <cr><lf>
Average of 3 phases.
Summary of Contents for BL3300 Series
Page 9: ...3 BL3300 SERIES MANUAL 01 28 2013 ALL MANUAL ADDENDUMS WILL FOLLOW THIS PAGE ...
Page 15: ...9 BL3300 SERIES MANUAL 01 28 2013 SECTION 2 UNPACKING AND INSTALLATION ...
Page 17: ...11 BL3300 SERIES MANUAL 01 28 2013 FIGURE 2 2 ...
Page 20: ...14 BL3300 SERIES MANUAL 01 28 2013 ...
Page 33: ...3 PHASE POWER CHASSIS MECHNICAL OUTLINE ...
Page 34: ...BL3300 CONTROL CHASSIS MECHANICAL OUTLINE ...
Page 35: ...BL33XX SERIES SINGLE PHASE POWER CHASSIS OUTLINE DRAWING 19 0 TYP 17 TYP 7 0 MAX ...
Page 54: ...6 1 SECTION 6 PARTS LIST AND DRAWINGS f ...
Page 56: ......
Page 57: ......
Page 58: ......
Page 60: ......
Page 62: ......