CALibrate:DATA <
value
>
Enters the calibration value read from a multimeter.
Parameter
Typical Return
Numeric value
(none)
Specify calibration value 0.0237: CAL:DATA 2.37E-2
The
must be ON to execute this command.
Do not use any unit suffix (such as A) in this command.
The CURRent calibration unit is always amps (DC) or amps-rms (AC).
The VOLTage calibration unit is always volts (DC) or volts-rms (AC).
CALibrate:DATE "<
date
>"
CALibrate:DATE?
Sets the calibration date. Use CALibrate:SAVE to store the date into nonvolatile memory.
Parameter
Typical Return
"<
date
>"
"2013 Jul 26"
Enter the calibration date: CAL:DATE "2014 Jul 26"
The
must be ON to execute this command.
This command increments the calibration count by 1.
The date may be up to 15 characters.
Use the format shown above (or the local language equivalent) to avoid confusion. For example, there
may be confusion as to whether "14/06/13" is June 14, 2013 or June 13, 2014. The format "2013 Jun
14" avoids that problem.
You may query the value regardless of whether the calibration state is on.
Keysight AC6800B Series Operating and Service Guide
117
3 SCPI Programming Reference
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...