ABORt Subsystem
ABORt commands cancel any triggered actions and returns the trigger system back to the Idle state.
Send
instead of ABORt to abort phase synchronization. ABORt commands are
also executed with the
command.
ABORt[:ALL]
Aborts both TRANsient and ACQuire operations.
Parameter
Typical Return
(none)
(none)
Abort TRANsient and ACQuire operations: ABOR:ALL
The trigger status at power-on is the same as that after an ABORt.
If an ABORt is sent when a transient or acquisition is in progress, the ACQuire group's acquired data
remains invalid.
If an ABORt is sent when the ACQuire group is not initiated and the acquired data that is held is valid,
the acquired data is retained.
ABORt:ACQuire
Aborts all ACQuire operations.
Parameter
Typical Return
(none)
(none)
Abort the triggered measurement: ABOR:ACQ
This command resets the WTG-meas and MEAS-active bits in the Operation Status registers.
If INIT:CONT:ACQ is ON, the ABORt:ACQuire command aborts the measurement but the instrument
immediately re-initiates new measurements.
ABORt:TRANsient
Aborts all TRANsient actions.
Parameter
Typical Return
(none)
(none)
Abort any TRANsient actions: ABOR:TRAN
This command resets the WTG-tran bit in the Operation Status registers.
Keysight AC6800B Series Operating and Service Guide
115
3 SCPI Programming Reference
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...