FREQuency Subsystem
FREQuency commands configure the instrument's output frequency.
[SOURce:]FREQuency[:CW] <
value
>|MINimum|MAXimum[,<
lower_limit
>,<
upper_limit
>]
[SOURce:]FREQuency[:CW]? [MINimum|MAXimum]
[SOURce:]FREQuency[:IMMediate] <
value
>|MINimum|MAXimum[,<
lower_limit
>,<
upper_
limit
>]
[SOURce:]FREQuency[:IMMediate]? [MINimum|MAXimum]
Sets the immediate AC frequency and optionally the frequency soft limits.
Parameter
Typical Return
40.0 to 500.0 Hz
+5.01
Set the output frequency to 50 Hz: FREQ 50
This command takes one or three arguments. You cannot have a low or high limit without the other.
This command applies to AC, AC+DC, EXT-DC, and EXT-AC.
The
default is 60 Hz.
You should set this value between the lower and upper frequency soft limits.
Frequency suffixes (HZ and KHZ) are allowed.
[SOURce:]FREQuency:LIMit:LOWer <
value
>|MINimum|MAXimum
[SOURce:]FREQuency:LIMit:LOWer? [MINimum|MAXimum]
[SOURce:]FREQuency:LIMit:UPPer <
value
>|MINimum|MAXimum
[SOURce:]FREQuency:LIMit:UPPer? [MINimum|MAXimum]
Sets the lower or upper AC frequency soft limit.
Parameter
Typical Return
40.0 to 500.0 Hz
Default 500 (UPPer) and 40 (LOWer)
+1.02
Set the frequency limits to 125 and 300 Hz:
FREQ:LIM:LOW 125
FREQ:LIM:UPP 300
FREQ:LIM:STAT ON
This command applies to AC, AC+DC, EXT-DC, and EXT-AC.
3 SCPI Programming Reference
130
Keysight AC6800B Series Operating and Service Guide
Summary of Contents for AC6801B
Page 1: ...Operating and Service Guide Keysight AC6800B Series Basic AC Sources...
Page 16: ...AC6802B 1 Getting Started 16 Keysight AC6800B Series Operating and Service Guide...
Page 17: ...AC6803B Keysight AC6800B Series Operating and Service Guide 17 1 Getting Started...
Page 18: ...AC6804B 1 Getting Started 18 Keysight AC6800B Series Operating and Service Guide...
Page 62: ......
Page 196: ...4 Calibration Verification and Service 196 Keysight AC6800B Series Operating and Service Guide...