DATA:POINts:EVENt:THReshold <
count
>
DATA:POINts:EVENt:THReshold?
Sets the number of measurements to store in reading memory before setting bit 9 in the Standard Oper-
ation Register group event register to 1. An event register is a read-only register that latches events from
the condition register. While an event bit is set, subsequent events corresponding to that bit are ignored.
For more information on the SCPI Status System, see
Parameter
Typical Return
34460A: 1 to 1,000 readings
34461A: 1 to 10,000 readings
34465A/70A: 1 to 50,000 readings (without the MEM option), or 1 to 2,000,000
readings (with the MEM option)
The default value is 1 for both models.
+125
Set the reading memory threshold to 125 measurements:
DATA:POIN:EVEN:THR 125
l
To report any subsequent events, the measurement count must first drop below the programmed
threshold before reaching the threshold again.
l
Use
to enable the Memory Threshold bit (bit 9 in the Standard Operation
Event register) to be reported to the Status Byte.
l
Once the Memory Threshold bit (bit 9 in the Standard Operation Event register) is set, it remains set
until cleared by
or
.
l
The Standard Operation Condition register shows the current condition. See
for further information.
l
This parameter is set to its default value after a Factory Reset (
) or Instrument Preset
(
).
Keysight True
volt
Series Operating and Service Guide