SYSTem:SECurity:IMMediate
(Requires the SEC option.) Sanitizes all user-accessible instrument memory except for the calibration con-
stants and reboots the instrument. This complies with requirements in chapter 8 of the National Industrial
Security Program Operating Manual (NISPOM).
Requires the optional security feature. See
for details.
The
NISPOM Sanitize
softkey and the SYSTem:SECurity:IMMEdiate command are equi-
valent. They are for customers, such as military contractors, who must comply with
NISPOM.
This feature destroys all user-defined state information, measurement data, and user-
defined I/O settings such as the IP address. This feature is not recommended for use in
routine applications because of the possibility of unintended data loss.
Parameter
Typical Return
(none)
(none)
Sanitize all user-accessible instrument memory:
SYST:SEC:IMM
l
Typically used before removing an instrument from a secure area.
l
Initializes all instrument settings to their Factory Reset (
) values. It also destroys all user-defined
state and preference information (see
MMEMory Subsystem - STATe and PREFerence Files
).
l
Increments the
SYSTem:TEMPerature?
Returns the instrument's internal temperature in °C.
Parameter
Typical Return
(none)
+2.8501
Return the instrument's internal temperature:
SYST:TEMP?
l
The value returned is not affected by
.
SCPI Programming Reference
402