[SENSe:]DATA2:CLEar[:IMMediate]
Clears the latest result(s) of the secondary measurement. This is typically used to clear the accumulated
history of the peak-to-peak function (
[SENSe:]VOLT:[DC:]SECondary "PTPeak"
or
).
Parameter
Typical Return
(none)
See example below
Configure DC voltage measurements using the 10 V range with 1 mV resolution. Select peak-to-peak as the
secondary measurement. Make and retrieve measurements.
CONF:VOLT:DC 10,0.001
VOLT:DC:SEC "PTP"
READ?;DATA2?
Typical Response: -1.84694879E-03;-2.23115575E-03,+5.98403324E-04,+2.82955909E-03
(Data
returned in this order: DCV reading, Min Peak, Max Peak, Peak-to-Peak)
Later, clear peak-to-peak secondary measurement data. Make and retrieve measurements.
DATA2:CLE
READ?;DATA2?
Typical Response: +1.29177114E-03;+6.07221096E-04,+3.43677658E-03,+2.82955548E-03
(Data
returned in this order: DCV reading, Min Peak, Max Peak, Peak-to-Peak)