Lake Shore Model 331 Temperature Controller User’s Manual
MNMX
Minimum and Maximum Input Function Parameter Command
Input:
MNMX <input>, <source>[term]
Format:
a,n
<input>
Specifies input to configure: A or B.
<source>
Specifies input data to process through max/min. Valid entries: 1 = kelvin,
2 = Celsius, 3 = sensor units, 4 = linear data.
Example: MNMX
B,3[term]
– Input B min/max function is on and processes data from the input sensor
units reading.
MNMX?
Minimum and Maximum Input Function Parameter Query
Input:
MNMX? <input>[term]
Format:
a
<input>
Specifies which input to query: A or B.
Returned:
<source>[term]
Format:
n
(Refer to command for description)
MNMXRST
Minimum and Maximum Function Reset Command
Input:
MNMXRST[term]
Remarks:
Resets the minimum and maximum data for all inputs.
MODE
Remote Interface Mode Command
Input:
MODE <mode>[term]
Format:
n
<mode>
0 = local, 1 = remote, 2 = remote with local lockout.
Example: MODE 2[term] –
Places the Model 331 into remote mode with local lockout.
MODE?
Remote Interface Mode Query
Input:
MODE?[term]
Returned:
<mode>[term]
Format:
n
(Refer to command for description)
MOUT
Control Loop Manual Heater Power (MHP) Output Command
Input:
MOUT <loop>, <value>[term]
Format:
n,
±
nnnnnn[term]
<loop>
Specifies loop to configure: 1 or 2.
<value>
Specifies value for manual output.
Example: MOUT
1,22.45[term]
– Control Loop 1 manual heater power output is 22.45%.
MOUT?
Control Loop Manual Heater Power (MHP) Output Query
Input:
MOUT? <loop>[term]
Format:
n
<loop>
Specifies which loop to query: 1 or 2.
Returned:
<value>
Format:
±
nnnnnn[term]
(Refer to command for description)
Remote Operation
6-35