C Commands
Lasair II Particle Counter Operations Manual
Page G-3
C Commands
Supported C Commands
C 0 yy
Set Date-Year. Two digit year
C 1 mm Set Date-Month
C 2 dd
Set Date-Day
C 3 hh
Set Time-Hour. Use 24 hour format.
C 4 mm Set Time-Minute
C 5 sec Set Time--Second
C 6 id_string Set Instrument ID
C 7 num Set Number of Samples
C 8 hh
Set Sample Time-Hour. Set sample time in order of hours, minutes and
seconds to avoid problems.
C 9 min
Set Sample Time-Minutes. Set sample time in order of hours, minutes
and seconds to avoid problems.
C 10 sec
Set Sample Time-Seconds. Set sample time in order of hours, minutes
and seconds to avoid problems. This command returns a Range Error if sec < 6, and
min = 0, and hours = 0.
C 11 [1/0] Set Sample Print on/off
C 12 hh
Set Delay Time--Hours. Set delay between samples.
C 13 min Set Delay Time--Minutes. Set delay between samples.
C 14 sec
Set Delay Time--Seconds. Set delay between samples.
C 16 [1/0]
Set continuous (repeat) mode on/off. When turned off, the number of
samples is set to 1.
C 21 chan name Set Analog ID Label. Set name for user analog channel 1-4
C 22 chan min
Set Analog 4 mA value. Set user analog channel minimum
engineering unit value.