SIGLENT
SSG5000X Programming Guide 103
3.8.3.3
Carrier Table
([:SOURce]:RADio:DMODulation:ARB:SETup:MCARrier:TABLe)
Command
Format
[:SOURce]:RADio:DMODulation:ARB:SETup:MCARrier:TABLe
INIT|APPend|<carrier_num>,”<waveform>”,<freq_offset>,<power>,<phase
>
[:SOURce]:RADio:DMODulation:ARB:SETup:MCARrier:TABLe? <pos>
Instructio
n
This command modifies the multicarrier table
Parameter
Type
String, String, Double ,Double ,Double
Parameter
Range
carrier_num: INIT | APPend | 1 ~ list rows,
waveform: None,
freq_offset: waveform sampling rate/4-maximum sampling rate/2 ~
maximum sampling rate/2- waveform sampling rate/4,
power: -40 ~ 0,
phase: -360 ~ 360,
pos: 1 ~ list rows.
Return
String, Double, Double, Double, Double, Integer
Default
SINE_WAVE,0,0,0,2MHz,200
Menu
ARB > Multi Carrier > Carrier Table
Example
:RADio:DMODulation:ARB:SETup:MCARrier:TABLe
INIT,SINE_WAVE,1000000,-10,20
:RADio:DMODulation:ARB:SETup:MCARrier:TABLe? 1
3.8.3.4
Carrier Save
([:SOURce]:RADio:DMODulation:ARB:SETup:MCARrier:STORe)
Command
Format
[:SOURce]:RADio:DMODulation:ARB:SETup:MCARrier:STORe
“<file_name>”
Instruction
This command stores the current multicarrier setup information
Parameter
Type
String