SIGLENT
SSG5000X Programming Guide 63
3.6.7.4
PM Deviation ([:SOURce]:PM:DEViation)
Command
Format
[:SOURce]:PM:DEViation
<value>
[:SOURce]:PM:DEViation
?
Instruction
Set the modulation deviation of the phase modulation (PM)
Get the modulation deviation of the phase modulation (PM)
Parameter
Type
Float, unit: rad
Parameter
Range
0.00001 rad ~ 5 rad
Return
Float, unit: rad
Default
1 rad
Menu
PM > PM Deviation
Example
:PM:DEViation 2
3.6.7.5
PM Rate ([:SOURce]:PM:FREQuency)
Command
Format
[:SOURce]:PM:FREQuency
<value>
[:SOURce]:PM:FREQuency
?
Instruction
Set the phase modulation (PM) frequency
Get the phase modulation (PM) frequency
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz, Default “Hz”
Parameter
Range
INTernal: SQUAre:0.01 Hz ~ 20 kHz SINE: 0.01 Hz ~ 100 kHz
INT+EXT: SQUAre:0.01 Hz ~ 20 kHz 0.01 Hz ~ 100 kHz
Return
Float, unit: Hz
Default
10 kHz
Menu
PM > PM Rate
Example
:PM:FREQuency 10 kHz
3.6.7.6
PM Sensitivity ([:SOURce]:PM:SENSitivity)
Command
Format
[:SOURce]:PM:SENSitivity
?
Instruction
Query the input sensitivity of the EXT MOD input in rad/v
Parameter None