SIGLENT
SSG5000X Programming Guide 77
Parameter
Range
UP|DOWN
Return
Enumeration
Default
UP
Menu
LF Sweep > Sweep Direction
Example
:LFOutput:SWEep:DIRect DOWN
3.6.10.3
Start Freq
([:SOURce]:LFOutput:SWEep:STARt:FREQuency)
Command
Format
[:SOURce]:LFOutput:SWEep:STARt:FREQuency <freq>
[:SOURce]:LFOutput:SWEep:STARt:FREQuency?
Instruction
Set the start frequency of sweep mode
Get the start frequency of sweep mode
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz, Default “Hz”
Parameter
Range
0.01 Hz ~ Stop frequency
Return
Float, unit: Hz
Default
500 Hz
Menu
LF Sweep > Start Freq
Example
:LFOutput:SWEep:STARt:FREQuency 100
3.6.10.4
Stop Freq
([:SOURce]:LFOutput:SWEep:STOP:FREQuency)
Command
Format
[:SOURce]:LFOutput:SWEep:STOP:FREQuency <freq>
[:SOURce]:LFOutput:SWEep:STOP:FREQuency?
Instruction
Set the stop frequency of sweep mode
Get the stop frequency of sweep mode
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz, Default “Hz”
Parameter
Range
Start frequency ~ Maximum frequency of LF
Return
Float, unit: Hz