SIGLENT
SSG5000X Programming Guide 123
Return
Float
Default
0
Menu
I/Q Control > I/Q Output > I/Q Output Atten
Example
:DM:IQADjustment:EXTernal:IQATten 1
3.8.6.11
I/Q Output Gain Balance
([:SOURce]:DM:IQADjustment:EXTernal:GAIN)
Command
Format
[:SOURce]:DM:IQADjustment:EXTernal:GAIN <val>
[:SOURce]:DM:IQADjustment:EXTernal:GAIN?
Instruction
This command sets the I/Q gain ratio for signals routed to the rear panel I
and Q output connectors
The variable <val> is expressed in units of decibels (dB)
Parameter
Type
Float
Parameter
Range
-4 ~ 4
Return
Float
Default
0
Menu
I/Q Control > I/Q Output > I/Q Output Gain Balance
Example
:DM:IQADjustment:EXTernal:GAIN 1
3.8.6.12
I Output Offset
([:SOURce]:DM:IQADjustment:EXTernal:DIOFfset)
Command
Format
[:SOURce]:DM:IQADjustment:EXTernal:DIOFfset <val>
[:SOURce]:DM:IQADjustment:EXTernal:DIOFfset?
Instruction
This command sets the differential offset voltage for an in-phase (I) signal
routed to the I output connectors
The variable <val> is expressed in units of volts (mV
–
V)
Parameter
Type
Float
Parameter
Range
-3 V ~ 3 V
Return
Float