SIGLENT
86 SSG5000X Programming Guide
Parameter
Range
RF|USER
Return
Enumeration
Default
RF
Menu
SENSOR > Frequency
Example
SENSe:SOURce RF
3.7.14
Frequency (:SENSe[:POWer]:FREQuency)
Command
Format
:SENSe[:POWer]:FREQuency <type>
:SENSe[:POWer]:FREQuency?
Instruction
Set the frequency for frequency type "
USER
"
Get the frequency for frequency type "
USER
"
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz, Default “Hz”
Parameter
Range
9 kHz ~ 3.2 GHz
Return
Float, unit: Hz
Default
None
Menu
SENSOR > Frequency
Example
SENSe:FREQuency 1 MHz
3.7.15
Level Offset State
(:SENSe[:POWer]:OFFSet:STATe)
Command
Format
:SENSe[:POWer]:OFFSet:STATe ON|OFF|0|1
:SENSe[:POWer]:OFFSet:STATe?
Instruction
Switch the power offset switch status
Get the power offset switch status
Parameter
Type
Boolean
Parameter
Range
ON|OFF|0|1
Return
Boolean
Default
0
Menu
SENSOR > Level Offset