Section 4: Pulse card concepts
Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
4-30
4200A-PMU-900-01 Rev. A December 2020
Example: pulse_meas_timing
This function sets the following pulse measure timing settings for five spot mean measurements for
channel 1 of PMU1:
pulse_meas_timing(PMU1, 1, 0.6, 0.8, 5);
Where:
•
Instr_id
= PMU1
•
chan
= 1 (channel 1)
•
StartPercent
= 0.6 (60%)
•
StopPercent
= 0.8 (80%)
•
NumPulses
= 5 (output one pulse)