Model 4200A-SCS Pulse Card (PGU and PMU) User's Manual
Section 4: Pulse card concepts
4200A-PMU-900-01 Rev. A December 2020
4-29
Example: pulse_meas_sm
This function sets channel 1 of the PMU for the spot mean discrete measure type to acquire the
voltage amplitude measurement, the current base level measurement, and the time stamps. It also
enables LLEC for amplitude:
pulse_meas_sm(PMU1, 1, 0, 1, 0, 1, 0, 1, 1);
Where:
•
Instr_id
= PMU1
•
chan
= 1 (channel 1)
•
AcquireType
= 0 (discrete)
•
AcquireMeasVAmpl
= 1 (enable)
•
AcquireMeasVBase
= 0 (disable)
•
AcquireMeasIAmpl
= 1 (enable)
•
AcquireMeasIBase
= 0 (disable)
•
AcquireTimeStamp
= 1 (enable)
•
LLEComp
= 1 (enable)
Example: pulse_meas_wfm
This function sets channel 1 of the PMU for the waveform discrete measure type to acquire the
voltage/current readings for the waveform and the time stamps. It also enables LLEC.
pulse_meas_wfm(PMU1, 1, 0, 1, 1, 1, 1);
Where:
•
Instr_id
= PMU1
•
chan
= 1 (channel 1)
•
AcquireType
= 0 (discrete)
•
AcquireMeasV
= 1 (enable)
•
AcquireMeasI
= 1 (enable)
•
AcquireTimeStamp
= 1 (enable)
•
LLEComp
= 1 (enable)