104
Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
:TIME <i,i,i>
Set the system time
:ERRor?
Returns error code, description
:KLOCk?
Returns if keylock state
:KLOCk ON | OFF | 1 | 0
Sets the keylock state
:COMMunicate
:USB?
Returns COMM
:SERial
:BAUD?
Returns the serial communication baud rate
:BAUD
Set the serial communication baud rate
:CONTrol?
Returns 0
:CONTrol
:RTS?
Returns 0
:RTS <n>
Accepts 0. All other values generate a parameter
error (only flow control none is supported).
:XONX?
Returns 0
:XONX
Accept 0. All other values generate a parameter
error (no handshaking modes supported).
:TYPE
:PARity?
Returns the serial communication parity bit set-
ting (NONE, EVEN, ODD)
:PARity <CCC>
Set the communication parity setting to NONE,
ODD, or EVEN
:GPIB
: SELF
:ADDRess?
Returns the GPIB address
:SET?
Returns VENT, 0.0
:SET
Only accepts Vent, 0. Not allowed to boot up a
Mensor product in any mode except vent mode
with a setpoint of 0 for safety reasons.
UNIT
:[PRESsure] <c>
Sets the pressure units. Accepts the unit code
number, or the string value representation, i.e.
psi.
See unit code table below.
:[PRESsure]?
Returns the pressure units string
:DEFine[N]?
Returns the user defined units name, multiplica-
tion factor. Index N is 1 or 2.