102
Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
:STATe ON | OFF | 1 | 0
Set the filter enable
:BAND?
Returns the filter window
:BAND <f>
Set the filter window
:FREQuency?
Returns the filter frequency
:FREQuency <f>
Set the filter frequency
:INLimits?
Returns current pressure reading (float), stable
flag (0 not stable, 1 stable)
:INLimits
:TIME?
Returns the stable delay in seconds
:TIME<f>
Set the stable delay in seconds
:RANGe?
Returns primary sensor maximum pressure,
units, and pressure type, i.e. “1000.000psig”
:RESolution?
Returns sensor resolution (integer 4 to 6)
:RESolution <n>
Set the sensor resolution
:SLEW?
Returns the rate in current units
SOURce
[:PRESsure]
[:LEVel]
[:IMMediate]
[:AMPLitude]?
Returns the setpoint in current units
[:AMPLitude] <f>
Sets the setpoint in current units
:VENT?
Returns 1 if in vent mode and pressure is not fully
vented (stable) yet, 0 otherwise. Proper usage:
put in vent mode, then query this value until it is 0
again to know the system is vented.
:VENT ON | OFF | 1 | 0 ON or 1 puts in vent mode , OFF or 0 puts in
Measure mode
:SLEW?
Returns the rate setpoint in current units
:SLEW <f>
Sets the rate setpoint in current units
:SLEW
:MODe?
Returns MAX when highspeed mode is active.
Returns LIN otherwise.
:MODe <CCC>
Set the control mode by putting in either High-
speed or Custom mode. Note: the overshoot flag
and control mode are not independent in Mensor
products.
:OVERshoot?
Returns 1 if the channel’s control behavior is >50.
Returns 0 if the control behavior <=50.