Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
Span?
<sp>XXXXXXX<cr><lf>
Returns span scale factor for active
transducer
Sparity
Even, ODD, NONE
Sets the serial and USB parity
Sparity?
<sp>CCCC<cr><lf>
Returns the serial and USB parity
Srqmask
Stable,Error or both
Sets the CPC6050 to issue a service
request(SRQ) over the IEEE when the
pressure control is stable, or an error
occurs. These are 80 hex and 40 hex
respectively.
Srqmask?
<sp>{string}<cr><lf>
Returns “stable”, ”error” or “error,
stable” depending on the SRQ.
Sstop
1 or 2
Sets the serial and USB stop bits
Sstop?
<sp>X<cr><lf>
Returns the serial and USB stop bits
Stable?
Returns YES if instrument is stable or
No
Stabledelay
0 to 3600
Sets the stable time to the number of
seconds specified
Stabledelay?
<sp>XXXXXXX<cr><lf>
Returns the stable time.
Stabletime
0 to 3600
Sets the stable time to the number of
seconds specified
Stabletime?
<sp>XXXXXXX<cr><lf>
Returns the stable time.
StableWin
%FS value
Sets the stable window as a %FS
StableWin?
<sp>n.nn<cr><lf>
Returns the stable window.
Standby
None
Instrument placed in Standby Mode
Standby?
<sp>(YES or NO)<cr><lf>
Returns yes if instrument is in Stand-
by, no if otherwise
Step
Value inside upper/lower limits and in-
side the range of the active transducer.
Sets the control step size for the
instrument
Step-
Jogs the setpoint down one step
Step+
Jogs the setpoint up one step
Step?
<sp>+n.nn<cr><lf>
Returns the control step for the instru-
ment
Step%
Value in % of current range
Sets the control step in % of current
range
Steppct
Value in % of current range
Sets the control step in % of current
range
Steppct?
<sp>n.nn<cr><lf>
Returns the current step in % of cur-
rent range
Supply_Detect
YES or NO
Detects whether enough supply
pressure is connected, see Section
6.4.3.10 Detection Flags
Supply_Detect?
<sp>(YES or NO)<cr><lf>
Returns whether detection is enabled
Switch_Chan
A, B
Sets the switch channel over the bus
and makes that channel the active
remote channel