Operating Instructions - CPC6050
81
Modular Pressure Controller
CPC6050
Measure?
<sp>(YES or NO)<cr><lf>
Returns YES if instrument is in mea-
sure. NO if otherwise
Meas_Reg
ON or OFF
Sets the regulation of the pressure
across the Measure/Control solenoid,
see Section 6.4.3.10 Detection Flags.
Meas_Reg?
<sp>(YES or NO)<cr><lf>
Returns whether measure regulation
is enabled
Mode
MEASURE, CONTROL, VENT
Sets the operation mode
Mode?
<sp>XXXXXX<cr><lf>
Returns the operation mode
Netmask
nnn.nnn.nnn.nnn
Sets the Ethernet network mask
Netmask?
<sp>nnn.nnn.nnn.nnn<cr><lf>
Gets the Ethernet network mask
OSversion?
<sp>n.n<cr><lf>
Returns the operating system version
Outform
1 to 7 – see table in Section 7.7.3
Sets the output format
Outform?
<sp>X<cr><lf>
Returns the output format – see table
in Section 7.7.3
Peakmax?
<sp>n.nn<cr><lf>
Returns the maximum pressure since
peakreset was sent
Peakmin?
<sp>n.nn<cr><lf>
Returns the minimum pressure since
peakreset was sent
Peakreset
None
Resets the peak values.
Port
nnnnnn
Sets the Ethernet port of the instru-
ment
Port?
<sp>nnnnn<cr><lf>
Returns the Ethernet port of the in-
strument
Precision
ON or OFF
Sets the precision control mode,
which sets the control behavior, stable
window, stable delay, and rate set-
point
Precision?
<sp>(YES or NO)<cr><lf>
Returns the status of the precision
mode
Ptype
Absolute or Gauge
Sets the instrument pressure type –
emulation only works if the optional
barometric transducer is installed. If
no barometer is present, standard
atmosphere is used for the barometric
reading when in emulation mode.
Ptype?
<sp>CCCCC<cr><lf>
Returns “Absolute” or “Gauge” for the
pressure type in native mode and
“Absolute Emulated” or “Gauge Emu-
lated” for the emulated pressure types
Purge
Value inside upper and lower limits,
never in the vacuum region
Start purge sequence with DUT pres-
sure as the setpoint. Only applicable
with option PURGE enabled