80
Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
Leak_Dwell?
<sp>n.nn<cr><lf>
Returns the dwell time in seconds
Leak_Pre_Dwell
0 to 3600
Pre-dwell parameter is the time in
between going to measure mode, and
the dwell time. Useful if there are
pressure transients after switching the
measure.
Leak_Pre_Dwell?
<sp>n.nn<cr><lf>
Returns the pre-dwell time.
Leak_Delta_Limit
Value within the range of the sensor
Threshold for a passing/failing leak
test.
Leak_Delta_Limit?
<sp>n.nn<cr><lf>
Returns the maximum pressure
change allowed before failure
Leak_Pass?
<sp>(YES or NO)<cr><lf>
Leak_Initial?
<sp>+n.nn<cr>
Returns the leak test initial pressure
Leak_Final?
<sp>+n.nn<cr>
Returns the leak test final pressure
Leak_Delta?
<sp>n.nn<cr><lf>
Returns the change in pressure since
start of test.
List?
<sp>Pri,1,2;Sec,1,2;Bar,1<cr><lf>
Returns list of available transducers
and turn downs for the active channel.
Listcal?
<sp>PRI,{sn},1,{mm
ddyyyy},2,{mmdd
yyyy};SEC,{sn},1,{mmdd
yyyy},2,{mmddyyyy}<cr>
<lf>
Returns the serial number of each
installed transducer and calibration
dates for each range
Listrange?
<sp>PRI,1,min,max,2,min,max;S
EC,1, min,max,2,min,max[;3RD,1,min,
max,2,min,max;4TH,1,min,max,2,min,
max]<cr><lf>
(bracketed single output only)
Returns the ranges of the installed
transducers for the active channel.
Listsensorinfo?
<sp>PRI,SN,PTYPE,ACCURACY,1,
min,max,2,min,max;SEC,SN,PTYPE,
ACCURACY,1,min,max,2,min,max[;3
RD,SN,PTYPE,ACCURACY,1,min,m
ax,2,min,max;4TH,SN,PTYPE,ACCU
RACY,1,min,max,2,min,max]<cr><lf>
(bracketed single output only)
Returns the serial number, pressure
type, accuracy, and ranges of the in-
stalled sensors for the active channel
Listconfig?
CHA, Regtype;PRI,1,min,max;SE
C,1,min,max; CHB, Regtype;PRI,1
,min,max;SEC,1,min,max; CHbaro,
Baro;Bar,min,max
Returns the regulator types (Regtype:
Pump, LPSVR, MPSVR, HPSVR,
EPSVR) along with all the ranges of
the installed transducers.
Localgravity
Value in ft/s^2
Sets the local gravity in feet/sec^2
Localgravity?
<sp>+n.nn<cr><lf>
Returns the local gravity in feet/sec^2
LowerLimit
Value inside primary xducer range in
current units.
Sets the lower control limit for the
instrument
LowerLimit?
<sp>n.nn<cr><lf>
Returns the lower control limit for the
instrument in current units.
Macaddress?
<sp>cc-cc-cc-cc-cc-cc<cr><lf>
Returns Ethernet MAC address
Measure
None
Instrument placed in Measure Mode