82
Operating Instructions - CPC6050
Modular Pressure Controller
CPC6050
Purge?
<sp>S,T<cr><lf>
Returns purge data where S repre-
sents state (0 = complete, 1 = local
purge, 2 = remote purge), T repre-
sents the estimated time to complete
in seconds
Purgeabort?
Aborts purge sequence if currently
running. Only applicable with option
PURGE enabled
RangeMax?
<sp>n.nn<cr><lf>
Returns the maximum range of the
active transducer in the current units.
RangeMin?
<sp>n.nn<cr><lf>
Returns the minimum range of the ac-
tive transducer in the current units.
Rate?
<sp>n.nn<cr><lf>
Returns the rate reading of the instru-
ment in current units/current time unit
(see Runits)
Rdecpt?
<sp>n<cr><lf>
Returns the number of rate decimal
points. (see Resolution)
Reference
{EXTVAC or ATM}
Sets the reference type for vacuum
pump connected to the reference, or
vented to atmosphere. This option is
available only on native gauge units.
Reference?
<sp>{EXTVAC or ATM}<cr><lf>
Returns the current reference type.
(see Reference)
Release?
15 char string.
EX:
Release? Test_stand_1
Returns:
<sp>(YES or NO), CCC…
CCC<cr><lf>
This command is used to release
control of the instrument in a multiple
computer environment.
Yes if release is successful
No if instrument is being controlled
with another computer
CCC… = name of controlling com-
puter or AVAILABLE
(see Acquire? and Unlock)
Reset_instrument_
time_on
Resets the instrument’s time on coun-
ter. See usage application for more
details
Reset_instrument_
control_time_on
Resets the instrument’s time on coun-
ter. See usage application for more
details
Resolution
<n>
Sets the number of significant digits.
(see decpt)
Resolution?
<sp>n<cr><lf>
Returns the number of significant
digits. (see decpt)
Rfilter
Value in %
Sets the % of the rate filter
Rfilter?
<sp>n.nn<cr><lf>
Returns the rate filter.
Rsetpt
Value in current units
Sets the rate setpoint
Rsetpt?
<sp>n.nn<cr><lf>
Returns the rate setpoint
Runits
Sec, min, hr
Sets the rate time unit