CALCulate:AVERage[:STATe]
CALCulate:AVERage[:STATe] <
enable
>
controls whether measurement
averaging is enabled (
ON
) or disabled (
OFF
).
Notes
1. When
CALCulate:AVERage[:STATe]
is
ON
, an individual channel number must
appear in a scan list only once, and use of
LISTL
in not allowed.
2. When
CALCulate:AVERage[:STATe]
is
ON
, channels must use manual ranging.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
enable
>
discrete (string)
ON | OFF
none
Comments
•
When averaging is on, each channel in the scan list(s) is measured a number of
times (as set by
CALCulate:AVERage:COUNt
). The average of those
measurements becomes the reading stored for that channel.
•
Related Commands:
CALCulate:AVERage:COUNt
,
CALCulate:AVERage[:STATe]?
•
*RST Condition:
CALCulate:AVERage:STATe OFF
Usage
CALC:AVER ON
Averaging for all scanned channels is ON.
CALCulate:AVERage[:STATe]?
CALCulate:AVERage[:STATe]?
returns a
1
if averaging is on, or a
0
if averaging
is off.
Comments
•
Returned Value: Numeric value either
1
or
0
. The C-SCPI type returned is
int16.
•
Related Commands:
CALCulate:AVERage[:STATe]
,
CALCulate:AVERage:COUNt
Usage
CALC:AVER?
A subsequent enter statement will return 0
or 1.
CALCulate
Chapter 5
HP E1313/E1413 Command Reference 165