The Questionable Data Group indicates when errors are causing lost or questionable
data. See Table 5-4 for the bit assignments.
Table 5-4. Questionable Data Group
Bit #
Dec Value Hex Value
Bit Name
Description
0 - 7
Not Used
8
256
0100
16
Calibration Lost
At *RST or Power-on Control Processor has found a
checksum error in the Calibration Constants. Read error(s)
with SYSTem:ERRor? and re-calibrate area(s) that lost
constants.
9
512
0200
16
Trigger Too Fast Scan not complete when another trigger event received.
10
1024
0400
16
FIFO Overflowed Attempt to store more than 65,024 readings in FIFO.
11
2048
0800
16
Over voltage
Detected on Input
If the input protection jumper has not been cut, the input
relays have been opened and *RST is required to reset the
module. Overvoltage will also generate an error.
12
4096
1000
16
VME Memory
Overflow
The number of readings taken exceeds VME memory space.
13
8192
2000
16
Setup Changed
Channel Calibration in doubt because SCP setup may have
changed since last *CAL? or CALibration:SETup command.
(*RST always sets this bit).
14-15
Not Used
STATus:QUEStionable:CONDition?
STATus:QUEStionable:CONDition?
returns the decimal weighted value of the
bits set in the Condition Register.
Comments
•
The Condition Register reflects the real-time state of the status signals. The
signals are not latched, and therefore, past events are not retained in this register
(see
STATus:QUEStionable[:EVENt]?
).
•
Returned Value: Decimal weighted sum of all set bits. The C-SCPI type
returned is uint16.
•
Related Commands:
CALibration:VALue:RESistance
,
CALibration:VALue:VOLTage
,
STATus:QUEStionable[:EVENt]?
,
STATus:QUEStionable:ENABle
,
STATus:QUEStionable:ENABle?
•
*RST Condition: No change.
Usage
STATUS:QUESTIONABLE:CONDITION?
Enter statement will return value from the
Condition Register.
STATus
Chapter 5
HP E1313/E1413 Command Reference 259