[SENSe:]DATA:FIFO:MODE
[SENSe:]DATA:FIFO:MODE <
mode
>
sets the mode of operation for the FIFO
reading memory.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
mode
>
discrete (string)
BLOCk | OVERwrite
none
Comments
•
In
BLOCk
(ing) mode, if the FIFO becomes full and measurements are still being
made, the new readings are discarded. In
OVERwrite
Mode, if the FIFO becomes
full and measurements are still being made, new readings overwrite the oldest
readings. In this situation, the FIFO contains only the latest 65,024 readings. In
both modes Error +3021, “ FIFO Overflow” is generated to let you know that
measurements have been lost.
•
This command will cause Error +3000, “ illegal while initiated” if trigger is initiated.
•
Related Commands:
[SENSe:]DATA:FIFO:MODE?
,
[SENSe:]DATA:FIFO[:ALL]?
,
[SENSe:]DATA:FIFO:HALF?
,
[SENSe:]DATA:FIFO:PART?
,
[SENSe:]DATA:FIFO:COUNt?
•
*RST Condition:
SENSe:DATA:FIFO:MODE BLOCk
Usage
SENS:DATA:FIFO:MODE OVERWRITE
Set FIFO to overwrite mode.
SENS:DATA:FIFO:MODE BLOCK
Set FIFO to block mode.
[SENSe:]DATA:FIFO:MODE?
[SENSe:]DATA:FIFO:MODE?
returns the currently set FIFO mode.
Comments
•
Returned Value: String value either
BLOCK
or
OVERWRITE
. The C-SCPI
type returned is string.
•
Related Commands:
SENSe:DATA:FIFO:MODE
Usage
SENSe:DATA:FIFO:MODE?
Enter statement returns either BLOCK or
OVERWRITE.