Example Command Sequence
ROUT:SEQ:DEF LIST<
n
>,(@101,103,105,107)
<n> can be 1 - 4.
SENS:FUNC:TEMP RTD,100,(@101,103)
100
Ω
RTD temperature channels
1 and 3.
SENS:FUNC:VOLT:DC (@105,107)
Autorange voltage channels 5
and 7.
ROUT:SCAN LIST1
Default is autorange voltage.
CALC:LIM:UPP:DATA 750,(@101,103)
Upper limit set to 750
°
C.
CALC:LIM:LOW:DATA 450,(@101,103)
Lower limit set to 450
°
C.
CALC:LIM:UPP:DATA .2,(@105,107)
Upper limit set to +.2VDC.
CALC:LIM:LOW:DATA -.01,(@105,107)
Lower limit set to -.01VDC.
CALC:LIM:STAT ON,(@101,103,105,107)
Enable overall limit testing.
CALC:LIM:UPP:STAT ON,(@101,103,105,107)
Enable upper limit testing.
CALC:LIM:LOW:STAT ON,(@101,103,105,107)
Enable lower limit testing.
SAMP:TIM LIST1,<
t
c
>
<t
c
> can be 10
µ
s to 32.768ms.
TRIG:SOUR TIM
TRIG:TIM:PER <
t
p
>
<t
p
> can be .1ms - 6.5536 s.
TRIG:COUN <
count
>
<count> can be 1 - 32768 or
INF.
ARM:SOUR <
source
>
<source> can be BUS, EXT,
HOLD, IMM, or TTLTrg0 -
TTLTrg7.
INIT:IMM
Starting
ARM event, either hardware or software
Acquisition starts
Stopping
Stops when Trigger Count is reached unless count is INF then use ABORt.
134 Understanding the HP E1313/E1413
Chapter 4