[SENSe:]FUNCtion:VOLTage[:DC]
[SENSe:]FUNCtion:VOLTage[:DC] [<
range
>,](@<
ch_list
>)
links the specified
channels to return DC voltage.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
range
>
numeric (float32)
.0625 | .25 | 1 | 4 | 16
(See Comments below)
VDC
<
ch_list
>
channel list (string)
100 - 163
none
Comments
•
The
<
range
>
parameter: The HP E1413 has five ranges: .0625 VDC, .25 VDC,
1 VDC, 4 VDC, and 16 VDC. To select a range, simply specify the range value
(for example, 4 selects the 4 VDC range). If you specify a value larger than one of
the first four ranges, the HP E1413 selects the next higher range (for example, 4.1
selects the 16 VDC range). Specifying a value larger than 16 generates Error -222,
“ Data out of range” . Specifying
0
selects the lowest range (.0625 VDC). Specifying
AUTO
selects autorange. The default range (no range parameter specified) is
autorange. Autorange is not allowed while
SENSe:FILTer:LPASs:STATe
is
ON
.
Error +3072, “ Autorange not allowed with SENSE:FILTER on” will be generated at
INITiate:IMMediate
time if the filter is
ON
and any channel specifies autorange.
<
range
>
may be specified in millivollts (mv).
•
If you are using amplifier SCPs, you should set them first and keep their settings
in mind when specifying a range setting. For instance, if your expected signal
voltage is to be approximately .1 VDC and the amplifier SCP for that channel has
a gain of 8, you must set
<
range
>
no lower than 1 VDC or an input out-of-range
condition will exist.
•
The
*CAL?
command calibrates channels based on Sense Amplifier SCP setup at
the time of execution. If SCP settings are changed, those channels are no longer
calibrated.
*CAL?
must be executed again.
•
This command will cause Error +3000, “ illegal while initiated” if trigger is initiated.
•
See “ Linking Channels to EU Conversion” in Chapter 3 for more information.
•
Related Commands:
*CAL?
,
INPut:GAIN:
…
commands
•
*RST Condition:
SENSe:FUNCtion:VOLTage AUTO,(@100:163)
Usage
SENSe:FUNC:VOLT (@140:163)
Channels 40 - 63 measure voltage in
autorange.
[SENSe:]
Chapter 5
HP E1313/E1413 Command Reference 245