Note
Executing
*TST?
returns the module to its
*RST
state.
*RST
causes the FIFO data
format to return to its default of ASC,7. If you want to read the FIFO for
*TST?
diagnostic information and you want that data in other than the ASCII,7 format, be
certain to set the data FIFO format to the desired format (
FORMat:DATA
command)
after completion of
*TST?
but before executing a
SENSe:DATA:FIFO:
…
query
command.
•
The C-SCPI type for this returned value is int16.
•
Following
*TST?
, the module is placed in the
*RST
state. This returns many of
the module’s programmed states to their defaults. See page 66 for a list of the
module’s default states.
•
*TST?
performs the following tests on the HP E1413 and installed Signal
Conditioning Plug-ons:
Digital Tests:
Test#
Description
1-3
Writes and reads patterns to registers via A16 and A24.
4-5
Checks FIFO and CVT.
6
Checks measurement complete (Measuring) status bit.
7
Checks operation of FIFO half and FIFO full IRQ generation.
8-9
Checks trigger operation.
Analog Front End Digital Tests:
Test#
Description
20
Checks that SCP ID makes sense.
30-32
Checks relay driver and FET multiplexer interface with EU CPU.
33,71
Checks opening of all relays on power down or input overvoltage.
34-37
Check FET multiplexer interface with A/D digital.
Common Command Reference
282 HP E1313/E1413 Command Reference
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com