DIAGnostic:CALibration:TARE[:OTDetect][:MODE]
DIAGnostic:CALibration:TARE[:OTDetect][:MODE] <
mode
>
sets whether Open
Transducer Detect current will be turned off (the default mode) or left on during the
CALibration:TARE
operation.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
mode
>
boolean (uint 16)
0 | 1
volts
Comments
•
When
<
mode
>
is set to
0
(the
*RST
Default), channels are tare calibrated with
their OTD current off. When
<
mode
>
is
1
, channels that have their OTD current
on (
DIAGnostic:OTDetect ON,(@<
ch_list
>
)) are tare calibrated with their OTD
current left on.
•
By default (
*RST
) the
CALibration:TARE?
command will calibrate all channels
with the OTD circuitry disabled. This is done for two reasons.
1. Most users do not leave OTD enabled while taking readings.
2. The
CALibration:TARE
operation takes much longer with OTD enabled.
However, for users who intend to make readings with OTD enabled, setting
DIAGnostic:CALibration:TARE:OTDetect:MODE 1
will force the
CALibration:TARE
command to perform the calibration with OTD enabled
(on channels specified by the user with
DIAGnostic:OTDetect:STATe
).
•
Related Commands:
CALibration:TARE
,
CALibration:TARE?
,
DIAGnostic:CALibration:TARE[:OTDetect][:MODE]?
,
DIAGnostic:OTDetect[:STATe]
•
*RST Condition:
DIAGnostic:CALibration:TARE:OTDetect:MODE 0
Usage
Configure OTD on during CAL:TARE.
DIAG:CAL:TARE:MODE 1
Set mode for OTD to stay on.
CAL:TARE?
Start channel tare cal.
DIAGnostic:CALibration:TARE[:OTDetect][:MODE]?
DIAGnostic:CALibration:TARE[:OTDetect][:MODE]?
returns the currently set
mode for controlling Open Transducer Detect current while performing
CALibration:TARE
operation.
Comments
•
Returns a
0
when OTD current will be turned off during
CALibration:TARE
.
Returns
1
when OTD current will be left on during
CALibration:TARE
operation.
The C-SCPI type returned is int16.
•
Related Commands:
DIAGnostic:CALibration:TARE[:OTDetect][:MODE]
,
DIAGnostic:OTDetect[:STATe]
,
CALibration:TARE?
•
*RST Condition:
DIAGnostic:CALibration:TARE:OTDetect:MODE 0
DIAGnostic