Once a channel’s autorange floor is set by
DIAGnostic:FLOor
, it remains until
reset by another
DIAGnostic:FLOor
command or by a
*RST
command.
•
A channel with an autorange floor can be manually ranged below the floor
(
[SENSe:]FUNCtion:
…
commands). When the channel is returned to autorange,
the autorange floor setting is still in effect.
•
Related Commands:
DIAGnostic:FLOor:DUMP
,
[SENSe:]FUNCtion:
…
commands.
•
*RST Condition:
DIAGnostic:FLOor 0,(@100:163)
Usage
DIAG:FLO 1,(@100,124)
Channel 0 and 24 can autorange no lower
than 1 V range.
DIAGnostic:FLOor:DUMP
DIAGnostic:FLOor:DUMP
sends the autorange floor settings for all 64 channels to
the FIFO.
Comments
•
The format of the values returned from the FIFO with a
SENSe:DATA?
command
depends on the format chosen by the
FORMat:DATA
command.
•
Related Commands:
DIAGnostic:FLOor[:CONFigure]
,
FORMat[:DATA]
,
[SENSe:]FUNCtion:
…
commands
•
*RST Condition:
DIAGnostic:FLOor:DUMP
places 0 values in the FIFO for all
64 channels.
DIAGnostic:INTerrupt[:LINe]
DIAGnostic:INTerrupt[:LINe] <
intr_line
>
sets the VXIbus interrupt line the
module will use.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
intr_line
>
numeric (int16)
0 through 7
none
Comments
•
Related Commands:
DIAGnostic:INTerrupt[:LINe]?
•
Power-on and *RST Condition:
DIAGnostic:INTerrupt:LINe 1
Usage
DIAG:INT:LIN 5
Module will interrupt on VXIbus interrupt
line 5.
DIAGnostic
Chapter 5
HP E1313/E1413 Command Reference 193